The Longfian Jay - Filter Changing is a stationary oxygen concentrator manufactured by Lonfian / Longfian. Recommended maintenance interval is 100hours. This service manual provides detailed technical information for healthcare professionals and biomedical technicians who service or operate the Longfian Jay - Filter Changing.
Manual Depth: 1/5 — Basic — basic maintenance only
| Flow Rate | N/A |
|---|---|
| Purity | N/A |
| Power | N/A |
| Weight | N/A |
| Outlet Pressure | N/A |
| Alarms | N/A |
| Maintenance | 100hours |
Indications of Clean or Replace Air Filter Foam JAY series There are two air filters in JAY serial models, first filter and the second one. Please see the following diagram1 and diagram4 for thetwo air filters. It is recommended that the first filter should be cleaned or replaced every 100hours, the second filters shall be cleaned or replaced every 500hours . I Fist filter 1.See Diagram 1 for the first filter Filters can be taken out and washed or replaced every 100 hours. Diagram 1 II Second filter 1.Open the top of storage box and take off the storage box by loosing the scrow. III Second filter 1.Find the second filter in in Diagram 4, there is an airway pipe connected to the third filter 1 Diagram 4 2.Pull off the airway tube from the air filter, then loose the filter chamber in counter…